Day-to-Day Operation
4
STERRAD
®
200 User’s Guide
33
•
Provide at least 25 mm (1 inch) of space between the electrode and the top of
the load.
No
t
e:
Do not stack instrument inside the trays. Do not stack
trays. Do not stack trays within trays. Do not wrap
instruments within the trays.
No
t
e:
If you are using rigid containers cleared by the FDA for
use in the STERRAD System, follow the same procedures
that are recommended for use with the STERRAD or
APTIMAX
®
Instrument Trays. Do not stack instruments
inside the containers. Do not stack containers. Do not
stack containers within containers. Do not wrap
instruments within the containers.
CAUTION: Metal objects must not come into contact with
the chamber walls, the doors or the electrode.
Contact with the walls, doors, or electrode could
damage the sterilizer or instruments.