PR-805/810/810
L
User Manual
58
PMT C
ALIBRATION
If you wish to modify the current Photometer correction factor, use the
PMT Calibration
feature to
perform this task. This is a local factor and does not permanently change any factory calibration factors
for the standard lens or any accessories. However, it will modify Photometer readings regardless of the
accessory in use
–
it is a global factor.
The factory default value for this factor is 1.00. To return the PR-8XX to its factory state, simply change
this factor back to 1.00 if it has been modified.
The Photometer factor can be modified using one of two methods. The user can enter a calibration
value, or enter the absolute and have the PR-8xx correct itself following a measurement.
Warning When measuring sources other than Illuminant A (CCT: 2856K) it is recommended
that the onetime PMT calibration option be used to calibrate the Photometer to the source
that is being measured. This will yield more accurate photometric results.
M
ODIFYING THE
F
ACTOR
M
ANUALLY
The user may manually enter a factor to modify the photometer correction factor.
Step 1)
Access the
PMT Calibration Screen
by clicking on
Menu
→
Setup
→
PMT Calibration.
FIGURE 61 - PMT CALIBRATION SCREEN
Step 2)
Click on the
Correction Factor
field. The following data entry screen appears:
Содержание Photo Research PR-810
Страница 2: ......
Страница 61: ...PR 805 810 810L User Manual 53 OVERVIEW Figure 56 IS 8xx Outer View...
Страница 77: ...PR 805 810 810L User Manual 69 Total power available Total power required FIGURE 74 USB POWER SCREEN...
Страница 88: ...PR 805 810 810L User Manual 80 FIGURE 96 BLUETOOTH CONNECTIVITY SETUP WINDOW...
Страница 120: ...PR 805 810 810L User Manual PR 8XX OUTLINE DRAWINGS 112...