SL
Analog Output (Optional)
Enable analog output. For analog output operations the shutter must be set to OPEN
following a measurement.
Syntax
: SLa[CR]
Where:
a = Enable/Disable Analog Output
0 = Disable
1 = Enable
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code) [CR][LF] (NNNN = Error code)
SN
Cycles to Average
Defines the number of measurements (cycles) to average. The range of cycles to average is
1 to 99. The default is 1.
Syntax
: SNaa[CR]
Where
: aa = Cycles to Average Range 1 to 99
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
SO
CIE Observer
Photometric and Colorimetric values can be calculated using either CIE 2
or 10
Standard
Observer data sets. Use this specifier to choose the CIE data set for calculations for the next
measurement. The default is 2
.
Syntax
: SOn[CR]
Where
: n = CIE Observer
2 = 2
10 = 10
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
SP
Primary Accessory
A Primary Accessory is one that replaces the standard objective lens (typically the MS-75)
during use and can be used in conjunction with an Add-on Accessory.
Syntax
: SPnn[CR]
Where
: nn = Accessory Code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: Accessory Codes can be found by running report 116 (command D116). See the Data
Codes section for specific details.
Содержание Photo Research PR-810
Страница 2: ......
Страница 61: ...PR 805 810 810L User Manual 53 OVERVIEW Figure 56 IS 8xx Outer View...
Страница 77: ...PR 805 810 810L User Manual 69 Total power available Total power required FIGURE 74 USB POWER SCREEN...
Страница 88: ...PR 805 810 810L User Manual 80 FIGURE 96 BLUETOOTH CONNECTIVITY SETUP WINDOW...
Страница 120: ...PR 805 810 810L User Manual PR 8XX OUTLINE DRAWINGS 112...