PR-805/810/810
L
User Manual
42
U
SING A N
O
BJE C TI VE
L
E N SE S
Aligning and focusing the Optical System
1. If using the MS-75, make sure the target is out of focus by defocusing the lens. If using a fixed
focus lens, such as the MS-2.5X, move either the target or instrument until the target is out of
focus.
2. Turn the diopter adjustment on the eyepiece until the measuring aperture (black spot in the
center of the field of view) is in sharp focus. If the entire aperture does not focus simultaneously,
make sure the
top
and
bottom
of the aperture are in focus.
3. Focus on the target. If using the MS-75 lens, rotate the lens focusing ring until the target is in
focus. If using a fixed focus lens, move either the instrument or the target until sharp focus is
achieved.
Do not use the eyepiece diopter adjustment to focus on the target.
Objective Lens
Diopter
adjustment
FIGURE 45 - PR-8XX/ PR-8XX FOCUSING
4. Make sure the measuring aperture falls within the lit area of the target. Failure to adhere to this
step will result in erroneous readings. If possible, or unless otherwise specified, we recommend
that the diameter of the aperture cover 50% to 80% of the smallest dimension of the object (e.g.
character
‘
1
’
in Figure 46).
Correct
Incorrect
FIGURE 46 - APERTURE ALIGNMENT
5. From the
Instrument Setup
menu, Clickthe
▲ ▼
icons adjacent to
Primary Accessory
until the
installed lens appears.
6. Make any other changes in the
Setup
menu such as
Sensitivity, Speed, Average
etc.
7. Click the
MEASURE button
to begin a measurement.
Содержание Photo Research PR-810
Страница 2: ......
Страница 61: ...PR 805 810 810L User Manual 53 OVERVIEW Figure 56 IS 8xx Outer View...
Страница 77: ...PR 805 810 810L User Manual 69 Total power available Total power required FIGURE 74 USB POWER SCREEN...
Страница 88: ...PR 805 810 810L User Manual 80 FIGURE 96 BLUETOOTH CONNECTIVITY SETUP WINDOW...
Страница 120: ...PR 805 810 810L User Manual PR 8XX OUTLINE DRAWINGS 112...