95
PR-805/810/810
L
User Manual
Setup
Command
Description
SE0
Select Max Exposure Time (PR-810L ONLY)
Enter the Max Exposure (Integration) time for the next measurement in milliseconds.
Possible values are 200
–
10,000 ms (.2 to 10 sec.). This setting only applies when Gain is set
to Maximum or Auto Gain.
Syntax
: SE0,ttttt[CR]
Where
: ttttt = exposure time in milliseconds
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
SF
Aperture Select
Select the aperture to be used for the next measurement.
Syntax
: SFa[CR]
Where
: a = aperture code
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Note: See Data Code 117 for details on aperture codes.
SG
Gain
Electronic gain modes, choices are x1, x10, x100, Maximum (PR-810L Only), Auto Gain.
Syntax
: SGg[CR]
Where
: g = Gain
g
Value
0
X1 (Default)
1
X10
2
X100
3
Maximum
99
Auto Gain
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
SK
User Sync Frequency
Enter the frequency (in Hertz) of the source being measured. The range is 20 to 400 Hz.
This command works in unison with the
SYNC Mode
setting. See the SS specifier for
complete details on setting the
SYNC Mode.
Syntax
: SKfff[CR]
Where
: fff = frequency in Hertz. Range is 20 to 400
Response
: 0000[CR][LF] If all OK, else
NNNN[CR][LF] (NNNN = Error code)
Содержание Photo Research PR-810
Страница 2: ......
Страница 61: ...PR 805 810 810L User Manual 53 OVERVIEW Figure 56 IS 8xx Outer View...
Страница 77: ...PR 805 810 810L User Manual 69 Total power available Total power required FIGURE 74 USB POWER SCREEN...
Страница 88: ...PR 805 810 810L User Manual 80 FIGURE 96 BLUETOOTH CONNECTIVITY SETUP WINDOW...
Страница 120: ...PR 805 810 810L User Manual PR 8XX OUTLINE DRAWINGS 112...