PR-805/810/810
L
User Manual
13
To use the Analog Output:
Step 1)
Connect the appropriate end of the cable to the instrument, and connect the other end
of the cable to a device (e.g. oscilloscope) with at least
1 MEGOHM (1,000,000 Ohms)
of
impedance to prevent loading down the output signal.
Step 2)
In the
Instrument Setup
screen, set
Adaptive
to
Disabled,
set
Gain
to
1X
, set
Analog
Output
to
Enabled
and
Measure Shutter
to
Always Open.
The
Frequency Response
can
be set to the u
ser’s
preference depending on the noise characteristics of the signal. In
OPEN,
the bandwidth is
20 KHz
(50 µsecs).
Step 3)
Set up the source and press the
Measure
button to begin the measurement.
FIGURE 14 - ANALOG OUTPUT INSTRUMENT SETUP.
Warning: If Adaptive mode is enabled after enabling Analog output, Gain and Frequency
Response values remain unchanged. Gain and Frequency Response (determined by
the adaptive algorithm) values will only change after a measurement is taken.