Sensor Installation
62
Modline 5
5.
Route the cable either through the center opening in the mounting plate or along the side walls
of the WJ-5 and out through the insulated opening in the back door.
Figure 54: Mounting Kit
3.13
Sensor Installation Checklist
To ensure accurate, reliable, and trouble-free operation, check the installation for the following:
Sensor is properly mounted and aligned with no obstructions in optical path.
Sensor lens is focused on target. If possible, target size is at least twice diameter of reticle size as seen in
the viewing sight (i.e. twice the calculated spot size at the viewing distance).
Rear Protective window has been securely installed on back of Sensor. After completing the sighting
and focusing routines, screw the protection window back onto the Sensor to maintain the NEMA 4
(IP65) environmental rating.
Sensor lens is protected by air purge and clean purge air if atmosphere is dirty, oily or corrosive. Do not
use purge air with oil or water. Use instrument air or filter adequately.
Sensors with the Dirty Window Detector Option require an APA air purge or other protection as
explained in Section 3.8
, page 34. If the purging air quality is poor, use
an efficient filter. In worst cases, leave the air purge assembled to the Sensor, but disconnect the air
supply.
Sensor is adequately protected by water or air cooling and/or additional insulation if ambient
temperature exceeds safe limits.
Sensor is not subjected to direct or reflected radiated heat from oven walls, flames, etc. that cause it to
MFL Mounting Flange Large Accessory
UAA Universal Adapter Accessory
Modline 5 Sensor (without DWD option)
Sensor Cable
Spacer Block
(Item 2)
Socket Head
Cap Screws (Qty 2)
(Item 4)
Mounting Plate (Item 3)
Lock washers (Qty 4)
(Item 6)
Hex Head Screws
(Qty 4) (Item 5)