Maintenance
Modline 5
123
Error Codes for Sensor diagnostic information are summarized as follows:
X101
Dirty Window Detection (or attenuation alarm if using a 5R series with the Attenuation
Alarm turned on)
X102
Out of calibration
X103
Detector block too hot
X104
Detector block too cold
X105
Sensor internal Case temperature too high
X106
Sensor internal case temperature too low
X107
Sensor failure: FAIL may be seen on the Sensor TEMP display
X108
Analog Output Current loop malfunction (open circuit)
X109
Dirty Window Detection malfunction
pINV
pINV is only displayed when the Invalid Alarm (iALM) function is set to ON and Peak
Picker or Track and Hold is On. This condition becomes an alarm and triggers the Alarm
relay and the ALARM word on the Sensor rear panel.
Temperature displayed is Peak Picked or Hold Mode Temperature values from prior
valid measurement conditions.
Environmental conditions can cause Sensor errors. High or low ambient
temperatures can cause Sensor case temperature too high or too low errors. Dirty lens
conditions can be avoided using air purging.
Investigate all alarms indicated by the ALARM word indicator. Temperature readings
on rear panel display, at Analog Output or from RS-485 Communications may be
inaccurate with Error Codes X101 through X109 alarms.
Ratio Temperature measurement conditions are invalid with INV and pINV alarms.
Содержание Modline 5
Страница 2: ......
Страница 4: ......
Страница 156: ...Notices 154 Rev L6 Feb 2017 Modline 5 9 Notices ...