Sensor Installation
54
Modline 5
3.11
Installation Drawings and Illustrations
3.11.1
Installation Drawing for the UAA – Universal Adapter for Tripod or Custom
Mounting
Group Mounting Option
Required Accessories. See Individual Accessory
Descriptions and Dimensions.
DWD
See Notes
1
¼-20 Bolt Pattern Mount
UAA
N. R.
2
Large Flange Interface Mount
UAA
MFL Add ESA (3)
3
Air Purge with ¼-20 Bolt Pattern Mount
UAA
APA
Add ESA (2)
4
Air Purge with Small Flange Mount
UAA
APA
MFS
Add ESA (2)
5
Air Purge with Large Flange Mount
UAA
APA
MFL Add ESA (2)
DWD is Dirty Window Detector Option. Refer to notes with grouping table in Section 3.10
. The ESA adds 1.82 in. (46.2 mm) to length of the assembled group.
Table 8: Recommended Accessory Groupings UAA
Figure 43: Front mounted
Figure 44: UAA with APA
Sensor front surface will be positioned at the back end of the APA or MFL. To determine working
distance (D) for spot size formula d=D/F, measure or calculate from that point.
When using the MFL (Group 2) or APA (Group 3, 4, 5), assemble these accessories to
the UAA first. Place Sensor fully into assembly and rotate Sensor to desired position.
Tighten clamping screw.
Содержание Modline 5
Страница 2: ......
Страница 4: ......
Страница 156: ...Notices 154 Rev L6 Feb 2017 Modline 5 9 Notices ...