Product Description
18
Modline 5
Series
Model
Sighting:
Visible or Laser
DWD None or
Installed
Transfer
Standard
Lens Type
(Two Characters) (Four Digits)
0 or 1
0 or 1
0 or T
(Two Characters)
Example and Explanation
5 R
1 4 1 0
1
1
0
RA Lens
5R
600 to 1400°C
D/100
Laser Sight
DWD Installed
This unit is not
a Transfer
Standard
Focusing Range:
13 inches to
infinity D/100
The Model Number explanation information provided above defines a Modline 5 Sensor having the
following characteristics and features:
Model 5R-1410 is a Series 5R Ratio Type Sensor with the temperature range of 600 to 1400°C (1112
to 2552°F) and D/100 Resolution.
The Sensor Sight is Laser through the lens (Visible sighting was not selected).
The Dirty Window Detector Option (DWD) is installed
The Sensor is not a Transfer Standard. (T = Transfer Standard)
The installed lens is for the 5R Series and provides the standard focusing range of 13 inches to
infinity. The Sensor Model and lens combination provide a D/100 Optical Resolution
2.5
Accessories
2.5.1
Interconnecting Cable – Required Accessory
An IRCON supplied interconnecting cable with the Sensor mating connector attached on end one and
termination prepared wires on the other end is required. Special assembly of the cable is required and
the cable is only sold with the Sensor connector attached.
Maximum cable length is 350 feet (107 meters). The cable length limitation is due to the power supply
voltage requirements of the Sensor. The power supply voltage drop across this length of cable is low
enough to provide proper Sensor operation. Power supply cable wires are 22 AWG (0.35 mm²). Input
and Output wires are 24 AWG (0.25 mm²).
RS-485, Alarm relay output, and analog outputs, can extend beyond the 107 meter limitation. See Section
4
, page 64 for more information.
Sensor Standard Teflon Interconnecting Cable
Maximum temperature: 200°C (392°F)
Maximum cable length: 107 meters (350 feet)
See Red Lion Bulletin PAX – P for completing the specifications for DPM.
The cable must be terminated into a Power Supply/Signal Interface Box (POI) or a
Terminal Strip Plate (TSP) to ensure that the Sensor and cable installation meets RF
immunity and emission standards for CE compliance. These items are described
below.