RS-485 Digital Communications
Modline 5
111
BT
– Report Isoblock Temperature (Read Only)
Returns the infrared detector isoblock temperature in °C.
Typical Value Range: 0 to 65
RC
– Instrument Temperature (Read Only)
Returns the unit’s internal case temperature in °F or °C.
Range: N/A
Expected value is within the Range: 0 to 65 °C.
VR
– Firmware Version (Read Only)
The VR command returns the Sensor’s firmware version number in the format: VV.RR.
VV is the major version number, RR is the revision number.
Range: N/A
TP
– Controller Type (Read Only) – Reserved for future use.
Always returns zero.
Range: Not Applicable
LS
– Laser Control
Command is available with Laser sighting option only. Turns the sighting laser on or off. Note that
the laser has an automatic timeout of twenty minutes.
Range: 0 or 1
Values: Turns on the laser (1) or turns off the laser (0).
The Laser will not energize during the System Check. Also, if the Laser is energized and a manual
(CHK function) or automatic (HOUR function) system check is initiated, the Laser will turn off.
Observe and follow cautions below and operating instructions in Section 5.11
, page 102 of this Modline 5 Manual.
Before turning on the Laser remotely with the DPM Digital Panel Meter Interface or
other devices using RS– 485 Communications, insure there are no personnel in the path
of the beam.
FT
– Features Matrix (Read Only)
Returns the features that this unit supports.
Note: When bits 1 and 2 are not set, Sensor is 52 Series Brightness Unit. Range: N/A
Bit Values: