
38
www.
inivos
.com
INSTRUCTION MANUAL
REVISION 4.0 • APRIL
2020
ISSUES
Issue
On-screen description
Explanation
Possible solutions and advice
27
Forced bottle door
Bottle door appears forced
28
Temperature Sensor Fault
Temperature sensor
disconnected/broken/not
powered
29
Humidity Sensor Fault
Humidity sensor
disconnected/broken/not
powered
30
Fluid calibration. Start
voltage too low.
Fluid sensor voltage not high
enough at start.
Check that the fluid level sensor is plugged in and that
there is fluid all the way up to the bottle cup.
31
Fluid calibration. Start
voltage not stable
Fluid sensor voltage not
stable at start.
Check that there are no leaks.
32
Failed to calibrate (drain)
tank in
Fluid sensor voltage didn’t
stabilise.
Check that the tank is empty. Did you connect a container
to empty the tank into? Has the pump failed or is it running
slowly? Are there any restrictions slowing the fluid?
33
Fluid calibration, voltage
when the tank empty too
high
Fluid sensor reads too high at
the end.
The tank should be empty. Did you connect a container to
empty the tank into? Has the pump failed or is it running
slowly?
34
Fluid calibration, range too
large
The di
ff
erence between full
and empty is too large.
Is there a leak in the pipe to the sensor (fluid in the pipe)?
35
Fluid calibration, range too
small
The di
ff
erence between full
and empty is too small.
Is there a leak in the pipe to the sensor (fluid in the pipe)?
36
Service error. Periodic
service required.
This ProXcide requires service
to operate safely and reliably.
37
Unable to register with
server
This could be because there
is no 3G signal
38
Uncontained air volume
ProXcide has injected ProXcin
into the treatment area,
but the humidity has not
increased significantly.
Allow this process to complete fully to the thumbs down
screen, so that you can safely enter the area. To make the
next process a success, please ensure that the room is
fully closed.
39
Float raised before start
There is a hardware issue
40
Low fluid
There isn’t enough fluid to
complete the run
Содержание ProXcide
Страница 8: ...8 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 1 2 ...
Страница 41: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 41 Issue 4 Pop up box for issue 4 ...
Страница 42: ...42 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 5 Pop up box for issue 5 ...
Страница 43: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 43 Issues 12 Pop up box for issue 12 ...
Страница 44: ...44 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 13 Pop up box for issue 13 ...
Страница 45: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 45 Issues 16 to 20 Pop up box for issues 16 20 ...