INSTRUCTION MANUAL
REVISION 4.0 • APRIL
2020
5.1.6
Plug the process monitor mains cable into a convenient electrical socket. Ensure that the
position of the process monitor and power adaptor wires does not cause a trip hazard or
obstruction outside the treatment area. Also ensure the controls and indicators on the monitor
are accessible and visible. You must only use Inivos approved power adapters and cable
extensions.
If significant risk is identified in positioning the monitor just outside of the treatment area
or if there is not an electrical socket available immediately outside the treatment area, then
two options exist. Either plug the process monitor power adaptor in inside the room being
treated and run the power lead under the door, or position the process monitor further afield
in a suitable and safe location where a socket is available, and as close to the treatment area
as possible. Under normal circumstances the wireless communication between the process
monitor and the main unit can reach up to 20 meters, however this depends on the signal
strength and can be a
ff
ected by the fabric of the building and the systems environment.
5.1.7
Extend and plug in the mains cable of the main ProXcide™ unit by grasping and pulling the
mains plug from the base of the main unit. An approved CE marked 13A extension lead may
be used if necessary. If significant risk is identified, it is recommended that the unit is plugged
in outside the room being treated with the power lead being run under the door. This allows
the unit to be fully powered down in the event of a building fire or emergency.
5.1.8
Switch the socket on. The ProXcide™ takes up to one minute to perform baseline checks
and load the system software – Top LED indicators will illuminate
white
. When the system is
booted up and ready to start the top LED indicators will illuminate
blue
.
Press the top button
on the main unit to connect the main unit to the process monitor. The LED indicators will
illuminate
green
upon successful connection, otherwise it will remain
blue
.
5.1.9
Present
and hold
your operator card when prompted by the screen following successful
connection of the main unit to the process monitor. This step protects against unauthorised
processes being started. Processes can only be started with an operator card which acts as
a unique key card. If the process monitor has been turned on without first turning on the main
unit, the screen shown below will appear on the process monitor.
5.1.10
Check the system fluid level, which is indicated immediately after switching the socket on.
The fluid level LED indicator is located next to the bottle door and will illuminate
red
if it
requires bottle replacement. The fluid level LED indicator will illuminate
blue
as it slowly fills
up. You must wait for the blue LEDs to reach the top (full). The LED indicators go up or down
depending on the fluid level in the main reservoir tank.
5.1.9
5.1.10
5.1.7
5.1
ProXcide™ Operator Instructions
Содержание ProXcide
Страница 8: ...8 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 1 2 ...
Страница 41: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 41 Issue 4 Pop up box for issue 4 ...
Страница 42: ...42 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 5 Pop up box for issue 5 ...
Страница 43: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 43 Issues 12 Pop up box for issue 12 ...
Страница 44: ...44 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 13 Pop up box for issue 13 ...
Страница 45: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 45 Issues 16 to 20 Pop up box for issues 16 20 ...