If the electrical supply to the unit fails during a decontamination process, it must be assumed
that the space still contains vapour and is dangerous to enter. Power failure will disrupt the
decontamination process and can occur in three ways:
1. Power to the main unit and the process monitor is lost:
Once power is resumed to both the
process monitor and the main unit, the ProXcide™ main unit will automatically go into deactivation
mode and the process monitor will display the machines deactivation status.
2. Power to the main unit only is lost:
The process monitor outside the treatment area will display
a timed message indicating communication to the main unit was lost. When power to the main unit
returns, it will automatically go into deactivation mode and the process monitor will reconnect and
display the updated system status.
5.4
ProXcide™ power or communication failure