APRIL 2020 • REVISION 4.0
INSTRUCTION MANUAL
13
The ProXcide™ uses high frequency ultrasonic
vaporisation technology combined with fans
to aerosolise microscopic droplets of ProXcin™
into the treatment space. The droplets rapidly
evaporate to create gaseous hydrogen peroxide
that is actively circulated to decontaminate the
exposed surfaces in the room.
The ProXcide™ auto-calibrates and self-monitors
ambient humidity in real-time throughout the
decontamination process to ensure the target
treatment parameters are maintained.
The ProXcide™ bio-decontamination system
is designed to deliver a log 6 reduction in
pathogens to all surfaces in the a
ff
ected area.
The treatment cycle is standardised to ensure a
consistent e
ff
icacy in all environments.
Section 3:
How the ProXcide
™
works
The entire procedure is managed remotely by
the process monitor, from activation through
to process completion. The process monitor
and main unit communicate using a built in
internal wireless system (that doesn’t require
hospital Wi-Fi to work). The touch-screen
process monitor enables full visibility of the
decontamination process at each stage
throughout the process.
As the process monitor is a wireless device, if
there are any issues with having the system
positioned directly outside of the treatment
space, the process monitor may be positioned
elsewhere, for example behind the nurses’
station or in an o
ff
ice.
The ProXcide™ can be terminated at any stage in
the process via the touchscreen by pressing the
red stop button. Throughout the setup phase
of the decontamination process, the process
monitor guides the operator through a series of
on-screen checks to ensure the process is set
up correctly and safely. Only once these checks
have been confirmed as complete can a process
be started.
At the end of the treatment phase, or after
manual termination of the process via the
stop button, the ProXcide™ will automatically
revert to its fail-safe deactivation mode in
which concentrations of hydrogen peroxide are
rapidly reduced. The treatment space must not
be entered during the deactivation process.
The time remaining is always indicated on the
process monitor.
3.1
Ultrasonic vaporisation technology
3.2
Remote activation and control
Содержание ProXcide
Страница 8: ...8 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 1 2 ...
Страница 41: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 41 Issue 4 Pop up box for issue 4 ...
Страница 42: ...42 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 5 Pop up box for issue 5 ...
Страница 43: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 43 Issues 12 Pop up box for issue 12 ...
Страница 44: ...44 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 13 Pop up box for issue 13 ...
Страница 45: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 45 Issues 16 to 20 Pop up box for issues 16 20 ...