36
www.
inivos
.com
INSTRUCTION MANUAL
REVISION 4.0 • APRIL
2020
ISSUES
Issue
On-Screen description
Explanation
Possible solutions and advice
5
The tank is not full.
- Did you insert a new full
bottle of ProXcin? If not,
please turn main unit o
ff
and
on again to that the bottle
door will open automatically.
DO NOT force the door
open. Replace bottle with
new full bottle. Remember
to push down firmly and
loosen cap.
The tank that holds the
ProXcin solution is not full
enough.
Make sure that a new full bottle of ProXcin is inserted
each time it is automatically requested. If you did not
insert a full bottle, turn the main unit o
ff
/on again and
follow user instructions to replace with a new full bottle. If
it was a new bottle, check:
•
If there are any leaks on the floor
•
If the bottle has been pushed down hard enough and
cap loosened to release the vacuum
8
Fluid sensor or pump issue
during active operation.
The fluid level is detected by
a sensor, and this may have
fallen. Or the pump may not
be working.
This would occur during active operation and therefore
H
2
O
2
is present in the room and you must wait for
deactivation to complete.
Check floor for leaks. If there are none, set-up and run the
process again - it may correct itself. If it fails again, contact
Inivos quoting Issue 8 for system swap and repair.
12
A sensor has failed.
It is safe to enter the room.
Please contact Inivos
quoting issue 12 to arrange
system swap while it is
repaired
This is a hardware issue with
one of the sensors.
This would occur when first powered on, therefore there is
no H
2
O
2
in the room and it is safe to enter. Please contact
Inivos quoting issue 12 to arrange system swap while it is
repaired by us.
13
Fluid sensor or pump
issue detected during
preparation phase.
Check floor for leaks. If
there are none, set-up and
run the process again - it
may correct itself. If it
fails again, contact Inivos
quoting Issue 13 for system
swap and repair.
The fluid sensor did not rise,
or the fluid was not pumped
into the tank in time.
This would occur during the preparation phase therefore
H
2
O
2
is not present and it is safe to enter the room.
Check floor for leaks. If there are none, set-up and run the
process again - it may correct itself. If it fails again, contact
Inivos quoting Issue 13 for system swap and repair.
14
There is a software issue.
An error has occurred in the
software program .
Please contact Inivos quoting issue 14 to arrange system
swap while it is repaired by us.
Содержание ProXcide
Страница 8: ...8 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 1 2 ...
Страница 41: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 41 Issue 4 Pop up box for issue 4 ...
Страница 42: ...42 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 5 Pop up box for issue 5 ...
Страница 43: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 43 Issues 12 Pop up box for issue 12 ...
Страница 44: ...44 www inivos com INSTRUCTION MANUAL REVISION 4 0 APRIL 2020 Issue 13 Pop up box for issue 13 ...
Страница 45: ...APRIL 2020 REVISION 4 0 INSTRUCTION MANUAL 45 Issues 16 to 20 Pop up box for issues 16 20 ...