W82
Calibration factor
too high!
If the calibration factor
was set to more than 5,
the calibration was not
successful. If a
calibration factor
between 2 and 5 was
confirmed, a warning
still occurs after
confirming this
successful calibration.
Check the chamber volume
settings and/or gas flow
The calibration leak rate
is smaller than entered.
Check the correct setting for
the calibration leak value!
Chamber volume and/or
gas flow are greater
than entered.
Check the chamber volume
settings and/or gas flow
W84
Calibration leak
signal too small
Calibration leak
defective
Check your calibration leak
Trigger value(s) too
high
Check the trigger
Volume/flow setting too
low
Check the accumulation
volume/the carrier gas flow
Measurement chamber
leaking
Check the measurement
chamber for leaks
W89
Overstep of the limit
value!
The device is
contaminated with
helium.
Do not switch off the device.
Leave the device running
while supplying fresh air to
the measurement and
reference lines until the
warning goes out.
If the warning occurs
frequently, increase the
contamination limit!
Calibration leak value
too high during
calibration
Use a smaller calibration leak
for the calibration!