5.4 Connecting to the power supply system
NOTICE
Material damage if power supply pack has the wrong specifications or is
connected improperly
A power supply pack that has the wrong specifications or is connected improperly can
destroy the unit.
► Use a suitable power supply unit, see below.
► If the short circuit current of the power supply unit is greater than 10 A, connect a
fuse between the power supply unit and the device.
► Use power supply cables with a sufficient cross-section, see below.
The following power supply unit types are suitable:
• Output voltage with electrically safe separation
• Output voltage: 24 V DC ± 16 %
• Current rating: min. 6 A. In general, the device consumes 70 W.
Cable length
Minimum diameter/cable type
8.5 m
0.75 mm
2
10 m
1 mm
2
/ AWG 18
15 m
AWG 16
17.5 m
1.5 mm
2
25 m
AWG 14
The diagram shows the pin assignment of the connector plug.
Connect the power cable to the device as described below.
Fig. 12: Power supply cable
1
Modify a cable to the required size with the supplied cable bushing.
2
Slide the bushing onto the connector plug and secure the connection with the
cap nut.
3
Connect the cable to the power supply.