5 | Installation
INFICON
36 / 90
T-Guard2_Operating-instructions-jina85en1-13-(2009)
5.9 Connecting measurement, reference and
vacuum lines
Use the filters supplied. They protect the lines from blockages and the sensor from
dust.
Measurement and reference lines cannot be shortened. If the length of the line or its
inner diameter is changed, the measurement speed and the measurement results are
affected. Only use the supplied lines, or those available as accessories.
► Push the hoses into the ports until they lock into position.
To release the lines, press the release ring and pull the line.
1
2
3
4
Fig. 16: Vacuum line connection
1
“IN” connection for the measuring line
2
“REF” (reference) connection for the reference line
3
“OUT” connection for the vacuum line (pump connection)
4
Release ring