10 | Warning and error messages
INFICON
78 / 90
T-Guard2_Operating-instructions-jina85en1-13-(2009)
If the message occurs
constantly during start-
up, the EEPROM on the
system board is
defective.
Contact INFICON customer
service.
W62
EEPROM
parameters lost!
A parameter was
changed and reset to
factory setting during
the software update.
The parameters
affected are listed in the
warning below.
Check the setting of the
modified parameter in the
corresponding software
menu and set the desired
parameter.
If the message occurs
constantly during start-
up, the EEPROM on the
system board is
defective.
Contact INFICON customer
service.
W64
There are warnings
pending!
Confirmed but not yet
valid warnings are
repeated every two
hours or at every new
start-up.
Check the warnings twice!
W65
Communication
error with fieldbus
module
Switch the device off and
back on. If the problem
persists, contact the
INFICON customer service.
W72
No communication
with analog card
Switch the device off and
back on. If the problem
persists, contact the
INFICON customer service.
W81
Calibration factor
too low!
If the calibration factor
was set to less than 0.2,
the calibration was not
successful. If a
calibration factor
between 0.2 and 0.5
was confirmed, a
warning still occurs after
confirming this
successful calibration.
The calibration leak rate
is larger than entered.
Check the correct setting for
the calibration leak value!