INV MEASURE High resistance means that the device is currently measuring. The
leak rate of the last measurement is measured if both this output
and the REC STROBE output are high resistance.
FINE/GROSS
High resistance means that the device is currently carrying out a
measurement cycle .
INV GROSS
LEAK
High resistance means that the highest trigger value has been
exceeded by at least a factor of 5.
INV
CONTAMIN.
High resistance means that the sensor has been contaminated with
too much helium.
INV
BACKGROUN
D
High resistance means that the sensor current displays a low-
helium state. This is a requirement for a sensitive measurement.