INFICON
Description | 4
T-Guard2_Operating-instructions-jina85en1-13-(2009)
21 / 90
4.4.3 Electrical data
Supply voltage
24 V DC ± 16 % / max. 6 A
Power consumption, typical
70 Watt
Protection class
IP 40
Overvoltage category
II
Thermal circuit breaker (resettable)
7 A, inert
4.4.4 Physical data
Max. inlet pressure
1300 mbar
Minimum detectable leak rate helium
1 x10
-6
mbar l/s
Maximum leak rate that can be displayed
15 mbar l/s
Measurement range
5 decades
Time constant of the leak rate signal (63
% of the final value)
< 1 s
Gas flow FINE
180 sccm
Gas flow GROSS
90 sccm
Helium sensor
Wise™ Technology
Max. carrier gas flow
1000000 sccm
Time until ready for operation
– 3…30 min, depending on the length of
the rest time
– up to several hours in the event of long
storage