Controlling the device using interfaces | 7
T-Guard2_Operating-instructions-jina85en1-13-(2009)
69 / 90
• In the “LR log” setting, the leak rate is calculated by:
leak rate = final value × 10
(U - 10) / volt / decade)
For example:
A voltage of 3.7 V provides a leak rate of 7.08 × 10
-5
mbar l/s for a final value of 10
-1
mbar l/s and 2 V per decade.
For “Current log.”, the same applies for current instead of leak rate.
• In the “LR log” setting, the leak rate is calculated by:
For example:
10 V correspond to the final value, e.g. 10
-3
mbar l/s, 3.7 V then correspond to e. g. 3.7
× 10
-4
mbar l/s.
For “Current lin.”, the same applies for current instead of leak rate.