3000 Micro GC Operating Manual
8.4 Consumable Replacement Procedures
8.4.1 Replacing the External 10 Micron Sample Inlet Filter Disk
1
Shut off any sample flow to the GC.
2
Wait for the 3000 Micro GC inlet to cool.
3
Disconnect any sample line or conditioner to the sample inlet filter.
4
Using a 5/16 in. wrench, remove
Part B
. See
Figure 8-36 Replacing the sample inlet filter disk
5
Remove the used filter disk from
Part A
.
6
Replace the used filter disk with a new one (PN 5183-4652).
7
Reassemble the sample inlet filter. Turn
Part B
clockwise until finger-tight, then
use a wrench to tighten an additional 1/4 turn.
CAUTION
Do not overtighten the sample inlet filter assembly to
3000 Micro GC. This will damage the dual-ended ferrule.
8
Attach sample line or conditioner to sample inlet filter.
9
Resume sample flow as needed.
Part B
Part A
Dual-end ferrule
Filter disk
To GC