3000 Micro GC Operating Manual
Chromatograms not
separating/no data is
showing
3000 Micro GC requires
power cycle
Restart 3000 Micro GC as instructed in
section 6.3.2, Restart, on page 6-8
.
View is set to
Overlay
In EZ IQ, click
View
>>
Tile All Data
to
generate separate chromatograms.
EZ IQ requires restart
Close EZ IQ and re-open. Restart the
computer if necessary.
Chromatograms not
separating/no data is
showing (continued)
Drivers are corrupt
Re-configure the drivers through EZ IQ
Config
.
Calibration is poor
Contamination in calibration
or carrier gas
Ensure the carrier gas purity is between
99.995% and 99.9995%.
Add hydrocarbon/moisture/oxygen traps or
particulate filters to the carrier gas line to
improve purity.
Ensure there are no contaminants in the
calibration gas.
Instrument requires power
cycle
Restart 3000 Micro GC as instructed in
section 6.3.2, Restart, on page 6-8
.
External leak
Check carrier gas lines using a leak detector.
Tighten all fittings.
Poor integration
Ensure all of the peaks of interest are
identified and fully integrated.
Backflush Time
set too
early (if applicable)
Increase the
Backflush Time
until all peaks of
interest appear on the chromatogram.
Table 9-1 Common chromatographic problems (continued)
Symptom
Possible Cause
Remedy
Содержание 3000 Micro GC
Страница 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...