5 - 13
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
5.1.5.3 Installation Instructions for the Pressure Reducer and Genie Filter Assembly
1
Turn off 3000 Micro GC and wait for the sample inlet fitting to cool.
WARNING
3000 Micro GC has inlets which can be heated to 140°C.
Touching the inlets at operating temperatures can result
in injury. Extreme care should be taken to avoid touching
the heated inlets.
2
Move the 3000 Micro GC handle fully away from the front cover.
3
Remove any fittings or hardware currently installed on the 3000 Micro GC
sample fitting.
4
Place a 1/4 in. male nut, back ferrule, and front ferrule onto one end of the outlet
tube provided. See
for ferrule orientation. See
tube orientation.
5
Loosely install a 5/16 in. nut, back ferrule, and front ferrule onto the sample inlet
fitting. Do not tighten. The outlet tube will be installed into this fitting later.
Figure 5-16 Installing ferrules
1/4 in. male nut
Back ferrule
Front ferrule
Back ferrule
Front ferrule
5/16 in. female nut
Содержание 3000 Micro GC
Страница 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...