4 - 13
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
4.2.5.2 Tools Required
Two 7/16 in. wrenches, for the sample inlet filter assembly
One 5/16 in. wrench, for user sample line to sample inlet filter
4.2.5.3 Sample Inlet Filter Assembly Installation
1
Shut off any sample flow to 3000 Micro GC.
2
Allow the sample inlet to cool.
3
Disconnect the sample line to the 3000 Micro GC input fitting using the 5/16 in.
wrench, while holding the sample inlet filter with a 7/16 in. wrench.
4
Inspect the sample inlet filter assembly and verify that the filter disk is in place.
If not, place a filter disk between parts A and B of the filter body and thread the
parts together until finger-tight. See
Figure 4-8 Exploded view of sample inlet filter assembly G2801-60900
5
Install the sample inlet filter assembly on the 3000 Micro GC inlet using the
dual-ended ferrule. Use the two 7/16 in. wrenches to tighten the filter halves
until snug.
5a
Tighten sample inlet filter assembly to 3000 Micro GC inlet an additional
1/4 turn using 7/16 in. wrench on filter Part A.
6
Connect the sample line to the male portion of filter Part B, shown in
while stabilizing the sample inlet filter assembly Part B using a 7/16 in. wrench.
CAUTION
Do not overtighten the sample inlet filter assembly to
3000 Micro GC. This will damage the dual-ended ferrule.
Part B
Part A
Filter disk
Dual-ended ferrule
To GC
Содержание 3000 Micro GC
Страница 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...