Remove the blue cap from the 1/8 in. NPT male input fitting and install a
user-provided 1/8 in. female NPT fitting and sample line. See
.
Figure 5-20 Installing the sample line
WARNING
If a hazardous sample is used, the sample exhaust must
be vented to a fume hood or other designated area for
hazardous waste disposal.
13
Connect a vent line to the vent fitting using standard 1/8 in. fittings. Run the
exhaust to a fume hood or other designated area for hazardous waste disposal.
Figure 5-21 Installing the vent tubing
PN G2816A
Tighten
Tighten
Flow adjustment knob
Remove blue cap
from NPT fitting and
install sample line
Install vent tube here
PN G2816A