5 - 6
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
8
While maintaining the alignment of the outlet tubing to the inlet fitting, tilt the
bracket up and snap it onto the front of the instrument. When properly mounted,
the bracket should be flat against the 3000 Micro GC frame. Insert the outlet
tubing into the 3000 Micro GC inlet fitting. See
Figure 5-5 Genie filter assembly, installed
NOTE:
The outlet tubing should insert easily into the 3000 Micro GC sample
inlet fitting. If not, slide the accessory along the 3000 Micro GC frame
until the outlet tubing is unstressed.
9
Tighten the 1/4 in. nut on the top of the Genie Filter Assembly.
10
Tighten the 5/16 in. nut on the 3000 Micro GC sample inlet fitting using the
appropriate size wrenches.
Inlet fitting
Sample outlet tube installed
G2817A on a 2-channel 3000 Micro GC shown
Содержание 3000 Micro GC
Страница 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...