3000 Micro GC Operating Manual
7.15 Apply the Valco Valve Program to EZ IQ Method
7.15.1 Valco Valve Setup in EZ IQ
1
In
EZ IQ
, click
Method >> Instrument Setup
.
2
Click the
Sampling
Figure 7-68 Instrument setup window - switching
3
Select the
Valve purge position:
from the drop-down list. This is the initial
valve position for purging before starting a single run in a sequence.
4
Type a
Purge delay time:
. This is the time defined for switching the valve to a
position which is specified in the
Valve purge position
after starting a run.
5
Save the method by clicking
File >> Method >> Save
.