3.3.3.4 Backflush Injector
The backflush assembly consists of:
a backflush injector
a precolumn
an analytical capillary column
The precolumn is a short section of column, whereas the analytical column is
usually longer. The precolumn and analytical column are connected in series.
The precolumn functions to prevent contaminants from entering the analytical
column. The sample first travels through the precolumn and then through the
analytical column. After components of interest enter the analytical column, the
flow is reversed through the precolumn, flushing the unwanted components to vent.
This
backflushing
of the precolumn prevents these contaminants from entering the
analytical column and causing deactivation or ghost peaks.
The backflush injector has two method parameters in EZ IQ software:
Backflush
Time
(in seconds) and
Inject Time
(in milliseconds). The
Backflush Time
parameter on the
Instrument Setup
window can be changed to suit analysis
needs. Valid settings for the
Backflush Time
are between 0-240 s. Recommended
settings for the
Backflush Time
are between 6-12 s. Valid settings for the
Inject
Time
are between 0-250 ms.
The backflush injector can also function in a semi-fixed mode. An
Inject Time
of
0 ms results in a fixed volume injected and is best used for component levels from
0.1% and higher. This will result in higher precision.
The backflush injector can function in a variable volume mode. An
Inject Time
greater than 10 ms will result in a higher injection volume. This used for component
levels below 0.1%.
Actual inject volumes are typically a few microliters of gas. For better sensitivity, set
a higher
Inject Time
, however, accuracy and precision will be negatively impacted.
Backflush times vary with column and setpoints and must be optimized for every
3000 Micro GC module.
Содержание 3000 Micro GC
Страница 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Страница 2: ......
Страница 6: ......
Страница 8: ......
Страница 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Страница 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...