4 - 2
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
4.1.3 Site Requirements
4.1.3.1 Electrical Requirements
A standard 100 to 240 V(ac) electrical outlet is required. A
Universal/Uninterruptable Power Supply (UPS) or surge protector is
recommended.
4.1.3.2 Space and Venting Requirements
For optimum 3000 Micro GC performance and lifetime, allow unrestricted airflow
around 3000 Micro GC to dissipate heat.
Safely vent sample streams—potentially toxic, noxious, or flammable
gases—outside the instrument and away from the operating area. Vent toxic gases
to a hood, chemical trap, or reaction medium.
Avoid venting the instrument into drafty areas, such as in front of a heating/cooling
vent.
4.1.4 computer Requirements
Microsoft Operating Systems
Windows XP Pro with Service Pack 3 or Windows 7
.Net 3.0 or 3.5 Service Pack 1
Dual Core 2 Ghz CPU, 2 GB RAM, >10 GB free space
DVD Drive
Network card for Ethernet connection to 3000 Micro GC (TCP/IP)