3000 Micro GC Operating Manual
6.2.2 Check the Duty Cycle
1
The duty cycle should be approximately 30 to 40% for column head pressure
(CHP) and Delta P (if fixed volume).
2
Is the Duty Cycle between 30 to 40%?
2a
YES—Process with analysis.
2b
NO—If the duty cycle is > 50%, check the carrier gas inlet pressure. Replace
the carrier gas filter (PN 3150-0602). Recheck the Duty cycle. If it is still > 50%,
contact
.
6.3 Utilities
The
Utilities
tab is on the 3000 Micro GC embedded web page. See
.
Figure 6-8 Utilities tab
6.3.1 Set Date/Time
In the
Utilities
tab, click
Set Date/Time
to set the correct date and time for the
3000 Micro GC.