95
Step
Instruction
9.
When the bump test is complete, the LCD returns to the main menu, and the green
LED is illuminated.
NOTE:
In the event of a failed sensor (Cl2, HCl, or NH3), the docking station ignores any large
reading that are above the sensor’s maximum reading. In addition, the docking station ignores
any negative readings whose absolute value is larger than the sensor’s maximum reading.
NOTE:
If the instrument fails the bump test, the IDS will perform a calibration, unless the
If the calibration or bump test fails, the
red LED illuminates, the IDS sounds an alarm, and the docking station will display the
appropriate “Bump Failure” or “Calibration Failure” message. The failed sensor type will be
displayed in parenthesis under the failure message.
5.5. Calibration
5.5.1. Forced Calibrations
You can force an IDS to calibrate an instrument by using the IDS menu. You may wish to do
this when you want to perform a calibration prior to the next calibration scheduled for the
instrument.
To force a calibration, follow the instructions listed below.
Step
Instruction
1.
Make sure that the IDS is on and that the proper gas cylinder is connected to the IDS
and that it is correctly configured in the DSSAC. See section 4.6 Configuring Gas
Cylinders for more information about configuring gas cylinders.
2.
Dock the instrument in the IDS.
3.
Press any of the keys on the IDS keypad to access the main menu.
NOTE:
The menu cannot be used when the IDS is performing an instrument action.
NOTE:
The menu cannot be used if the
Menu Locked
setting for the IDS in DSSAC is set to
“Yes.”
Step
Instruction
4.
Use the ARROW keys on the keypad until Instrument is highlighted.
5.
Press the ENTER key. The Instrument menu appears.
6.
Use the ARROW keys on the keypad to highlight Calibrate. Press the ENTER key.
Содержание DSX-L
Страница 1: ...Start up Guide DSX L Local Server Mode Edition 9 January 24 2020 Part Number 17156008 ...
Страница 4: ...iv ...
Страница 8: ......
Страница 14: ......
Страница 21: ...17 3 For Installation Type choose Role based or feature based installation and click Next ...
Страница 24: ...20 ...
Страница 44: ...40 Step Instruction Figure 3 14 Entering the System Administrator Password 3 Start installation ...
Страница 64: ...60 13 Verify SQL has been added to the Exceptions list and click OK ...
Страница 83: ...79 Figure 4 11 Adding Gas from Industrial Scientific ...