94
LCD Display
LED
Alarm
Description
Unavailable Instrument Red
On
There is an unidentified problem with the
instrument.
Undocked Instrument
Red
On
An instrument was undocked from the IDS
during an event.
Unsupported Software
Version
Red
On
The docked instrument’s software is not
supported for use with the docking station.
5.4. Forced Bump Tests
You can force an IDS to run a bump test on an instrument by using the IDS menu. You may
wish to do this when you want to run a bump test before the next automatic bump test scheduled
for the instrument.
To force a bump test, follow the instructions listed below.
Step
Instruction
1.
Make sure that the IDS is on and that the proper gas cylinder is connected to the
Docking Station and that it is correctly configured in the DSSAC. See the
Configuring Gas Cylinders section for more information about configuring gas
cylinders.
2.
Dock the instrument in the IDS.
3.
Press any of the keys on the IDS keypad to access the main menu.
NOTE:
The menu cannot be used when the IDS is performing an instrument action.
NOTE:
The menu cannot be used if the
Menu Locked
setting for the IDS in DSSAC is set to
“Yes.”
Step
Instruction
4.
Use the ARROW keys on the keypad until Instrument is highlighted.
5.
Press the ENTER key. The Instrument menu appears.
6.
Use the ARROW keys on the keypad to select Bump Test. Press the ENTER key.
7.
The LCD displays a confirmation prompt: “Are you sure?” Use the ARROW keys
on the keypad to highlight Yes and then press the ENTER key. If you select No, the
LCD returns to the main menu.
8.
The IDS begins performing the bump test. The LCD displays “Testing Instrument.”
The yellow LED is illuminated.
Содержание DSX-L
Страница 1: ...Start up Guide DSX L Local Server Mode Edition 9 January 24 2020 Part Number 17156008 ...
Страница 4: ...iv ...
Страница 8: ......
Страница 14: ......
Страница 21: ...17 3 For Installation Type choose Role based or feature based installation and click Next ...
Страница 24: ...20 ...
Страница 44: ...40 Step Instruction Figure 3 14 Entering the System Administrator Password 3 Start installation ...
Страница 64: ...60 13 Verify SQL has been added to the Exceptions list and click OK ...
Страница 83: ...79 Figure 4 11 Adding Gas from Industrial Scientific ...