cBot 2 System Guide
43
Perform a Post-Run Wash
1
Wash the plate on the thermal stage with deionized water to remove any salts. Dry
with a lint-free cleaning tissue.
2
Fill the wash reservoir with approximately 12 ml deionized water and close the
instrument lid.
3
Select the checkbox to indicate that water is present, and then select
Wash
.
4
When the wash is complete, blot any excess water remaining in the wash reservoir.
Avoid the outlet ports to prevent fibers from entering the holes.
5
Select the checkbox to indicate that the wash reservoir is dry, and then select
Exit
.
The Start screen opens and your cBot 2 is ready for another run.