P
re
p
a
ri
n
g
R
e
a
g
e
n
ts
26
Material # 20004364
Document # 15065681 v01
Prepare the ExAmp Reaction
Prepare the ExAmp reaction master mix immediately before use.
1
Invert EPX1 and EPX2 to mix.
2
Briefly centrifuge EPX1, EPX2, and EPX3.
3
Prepare the ExAmp master mix in a 1.5 ml tube as follows.
a
Add 210 µl EPX1.
b
Add 30 µl EPX2. Slowly pipette to mix.
c
Add 110 µl EPX3. Slowly pipette to mix. Make sure that the bottom of the tube is
free of bubbles.
4
Add 35 µl master mix to the bottom of each well of the 8-tube strip. Slowly pipette to
mix. Change tips between samples.
5
Cap the tubes and briefly centrifuge.
6
Set aside on ice until you are ready to load the cBot.
CAUTION
Load the 8-tube strip onto the cBot within
15 minutes
.