cBot 2 System Guide
27
HiSeq High Output Flow Cell
To prepare reagents, thaw and inspect the reagent plate. The reagent plate takes
approximately 60 minutes to thaw using a room temperature water bath. Alternatively, you
can thaw reagents at 2°C to 8°C overnight, not to exceed 16 hours.
About Reagents
}
When vortexing or inverting the cBot reagent plate, keep a hand on top of the tubes.
Thaw the cBot Reagent Plate
1
Remove the cBot reagent plate from -25°C to -15°C storage.
2
Make sure that the tube strips are securely seated in the plate.
3
Thaw in a room temperature water bath for 60 minutes.
Prepare the cBot Reagent Plate
1
Invert the cBot plate to mix reagents.
2
Vortex to dislodge trapped air bubbles.
3
Tap on a hard surface to collect reagent droplets at the bottom of the tubes.
Alternatively, pulse centrifuge.
4
Make sure that the tubes are free of air bubbles, securely seated, and ordered as
numbered.
5
Promptly
proceed to setting up the cBot.
6
If you are sequencing Nextera libraries on a TruSeq v3 flow cell, proceed to
before setting up the cBot.
Prepare HP10 (TruSeq v3)
Prepare HP10 for use on the cBot only when using Nextera libraries on a TruSeq v3 flow
cell. HP10 is also compatible with other Illumina library types.
1
Remove HP10 from -25°C to -15°C storage.
2
Thaw in a beaker of room temperature deionized water for 20 minutes.
3
Add 150 µl HP10 to each tube of an 8-tube strip.
4
Set aside on ice.
5
Promptly
proceed to setting up the cBot.
Содержание cBot
Страница 4: ...iv Material 20004364 Document 15065681 v01 ...
Страница 16: ...10 Material 20004364 Document 15065681 v01 ...
Страница 22: ...16 Material 20004364 Document 15065681 v01 ...
Страница 66: ...60 Material 20004364 Document 15065681 v01 ...
Страница 76: ...70 Material 20004364 Document 15065681 v01 ...
Страница 88: ...Index 82 Material 20004364 Document 15065681 v01 ...