5-3
5
5.4
SELF TEST ERROR CODES
At power turn on, the 9099 conducts a selftest of its major components. The
test takes about 5 seconds. During the selftest the PWR LED is on and the
RDY LED is off. The network LAN and ACT LEDs may be on during selftest.
A successful test ends when the RDY LED on. Test failures are indicated by
the blinking LED patterns shown in Table 5-1. If a self test failure occurs, turn
the unit off for 10 seconds and turn power back on. If the failure persists, refer
to paragraph 5.7 for repair instructions. Note that some of the failures could
occur while the 9099 is running.
TAbLE 5-1 9099 SELF TEST ERROR CODES
Card LEDs
RDY LAN ACT RDY TALK LSTN SQR ERR Fault
⊕
-
-
-
-
-
-
-
fatal error (CPU,
FLASH,
RAM..etc.)
⊕
-
-
-
-
-
-
-
fatal error (power
supply)
⊕
-
-
-
-
-
-
Β
LAN IC, Network
Socket Failure or
DHCP. See 5.5
⊕
-
-
-
-
-
Β
-
GPIB IC or GPIB
Transceivers
⊕
-
-
-
-
-
Β
Β
Configuration Error
or Flash Failure
⊕
-
-
-
-
Β
-
-
OS Issued Exit
⊕
-
-
-
-
Β
-
Β
RAM IC or Memory
overflow error
⊕
-
-
-
-
Β
Β
-
OS Error
⊕
-
-
-
-
Β
Β
Β
Flash Error
Symbols:
⊕ =
solid on, B = blinking