1-4
1
1.4.
VXI-11 CONFORMANCE
The 9099 complies with the VXI-11.3 Specification.
1.4.1 RPC Protocol
The RPC protocol conforms to ONC RPC Version 2.
1.4.2 Sockets
The 9099's VXI-11 service supports 15 TCP/IP sockets for client communica-
tion with a maximum of 4 active sockets. The sockets are normally opened and
closed by the clients. The unit will close the socket and release all resources
if a broken connection is detected or when the link count goes to zero if Auto
Disconnect is enabled.
There is a separate socket for UDP RPC Port Mapper communication.
1.4.3 Channels
Supports Core, Abort and Interrupt channels. Core and Abort channels each use
a socket connection. A reverse Interrupt channel is a TCP/IP socket connection
that does not count against the 15 client communication sockets limit.
1.4.4 Device
Links and Locks
The 9099 supports a maximum of 64 device links and 64 locks that can be used
over multiple Core channels by one or more clients.
1.4.5 VXI-11
Interface Name
The 9099 has only one instrument personality and the default name is inst0.
The name may be changed to any 8 character string.
1.4.6 VXI-11.3
Supported Functions
The 9099 supports all VXI-11.3 functions including:
create_link
destroy_link
create_intr_channel destroy_intr_channel
device_lock device_unlock device_abort
device_read
device_write
device_clear
device_trigger
device_remote device_local device_readstb create_intr_channel
device_intr_SRQ device_enable_SRQ
Содержание 9099
Страница 20: ...1 16 1 Figure 1 1 9009 Outline Dimensions ...
Страница 24: ...1 20 1 Figure 1 3 9099 Certificate of Compliance ...
Страница 54: ...3 8 3 Figure 3 1 Status Reporting Structure Flash Configuration Reset ...
Страница 94: ...3 48 3 This page left intentionally blank ...
Страница 100: ...4 4 6 This page is left intentionally blank ...
Страница 161: ...A 49 A4 Refer to the ErrorLogger utility for the error value definitions Error Meaning 0 No error ...