3-24
3
Keyword
Default
Description
Value
TAbLE 3-3 SCPI COMMANDS AND QUERIES
(CONTINUED)
a decimal value equal to the binary sum of the
data. Multiple words are separated by commas.
HEXL converts each four bit nibble into the ASCII
characters 0-9 and A-F. TABLE allows the user
to define his own character set. All talk strings
end with a linefeed. Values are ASCii | HEXL |.
i.e.
ASCii example = 128,5,255
HEXL example = 8000, 05FF
CALibrate
Starts the Calibrate Branch
:IDN <string>
blank
Sets the user IDN message. String is up to 72
characters and consists of four comma separated
fields (manufacturer, model code, serial number
and product revision) . e.g. ICS Electronics, 9099,
S/N 1509125, Rev 00.05 Ver 15.09.04 The word
'Model' may not be used in the IDN string. S/N is
a non-case sensitive serial number identifier and
must be present to display the user's serial number
on the Welcome page.
:DATe <date>
00/00/0000 Saves the date and the current configuration. date
is a 10 character string.
:DATe?
Queries the calibration date. The response is
00/00/00 when the unit is not calibrated. Enter a
valid date to indicate when the unit is calibrated
or configured.
DEFault
Resets the unit to its factory settings and resets
CAL:DATE to 00/00/0000. The network settings
and user IDN string are not reset. Do a CAL:DATE
command to set the calibration date.
:LOCK
Off
Disables the lockable configuration commands
listed in Table 1-3 when On. Locked commands are
not visible to the user. Value is
1(On) | 0(Off).
Содержание 9099
Страница 20: ...1 16 1 Figure 1 1 9009 Outline Dimensions ...
Страница 24: ...1 20 1 Figure 1 3 9099 Certificate of Compliance ...
Страница 54: ...3 8 3 Figure 3 1 Status Reporting Structure Flash Configuration Reset ...
Страница 94: ...3 48 3 This page left intentionally blank ...
Страница 100: ...4 4 6 This page is left intentionally blank ...
Страница 161: ...A 49 A4 Refer to the ErrorLogger utility for the error value definitions Error Meaning 0 No error ...