3-44
3
TAbLE 3-6 ERRORLOG ERROR CODES
Error Code Error Description
1
VXI-11 Syntax Error
3
GPIB Device Not Accessible
4
Invalid VXI-11 Link ID
5
VXI-11 Parameter Error
6
Invalid VXI-11 Channel. Channel Not Established
8
Invalid VXI-11 Operation
9
Insufficient Resources (normally related to Link IDs or Locks)
11
Device Locked By Another Link ID
12
Device Not Locked
15
I/O Timeout Error
17
I/O Error
21
Invalid GPIB Address Specified
23
Operation Aborted (indicator, not a true error)
29
Channel Already Established
60
Channel Not Active
110
Device Already Locked
111
Timeout Attempting To Gain A Device Lock
999
Unspecified/unknown error
1000
VXI-11 Protocol Error
2000
RPC Protocol Error
2001
Unsupported RPC Function
2002
Insufficient RPC Message Length
NOTE: Usage of the Agilent I/O Library may result in some error log entries. Some
of these errors occur when the IO Library tries to determine if the VXI-11 device is an
Agilent instrument. Others are due to the incorrect usage of the VXI-11 protocol in the
current revision of the library. Agilent is aware of these protocol errors and will correct
these errors in later releases of the Agilent I/O Suite. These errors normally only show
up during the opening of a SICL/VISA device and can be safely ignored
.
The occurrence of a soft error will cause the ErrorLog utility to generate a
single-line error report. It consists of a time/date that the error was detected,
the numeric code of the error and a short English translation of the error code.
The Error Log Error Codes are listed in Table 3-6.
Normal usage of the ErrorLog utility can provide two key pieces of informa-
tion. The first is that an error did happen. Normal operation would not result in
errors and the occurrence of an error can indicate an abnormal condition that
should be investigated. Therefore it may be advantageous to have the ErrorLog
utility operating over a period of time if unexpected errors occasionally happen
and/or a new client utility is being developed.
Содержание 9099
Страница 20: ...1 16 1 Figure 1 1 9009 Outline Dimensions ...
Страница 24: ...1 20 1 Figure 1 3 9099 Certificate of Compliance ...
Страница 54: ...3 8 3 Figure 3 1 Status Reporting Structure Flash Configuration Reset ...
Страница 94: ...3 48 3 This page left intentionally blank ...
Страница 100: ...4 4 6 This page is left intentionally blank ...
Страница 161: ...A 49 A4 Refer to the ErrorLogger utility for the error value definitions Error Meaning 0 No error ...