30
ENS 3000 IO-Link
Edition: 2020-04-20
HYDAC ELECTRONIC GMBH
Part No.: 669898
18 Technical
data
Note: Reverse polarity protection of the supply voltage, excess voltage, override and short circuit protection are
provided.
FS
(
F
ull
S
cale) = relative to complete measuring range
1)
Other fluids on request
2)
Observe ambient temperature range
3)
Specified at resting level
4)
Environmental conditions according to 1.4.2 UL 61010-1; C22.2 No. 61010-1
5) with mounted mating connector in corresponding protection type
Input data
Measurement range
mm
(inch)
170
(6.70)
290
(11.40)
390
(15.35)
590
(23.20)
Rod length
mm
(inch)
250
(9.80)
410
(16.20)
520
(20.50)
730
(28.70)
Max. speed of the
fluid level change
mm/s
(inch/s)
40
(1.57)
60
(2.36)
80
(3.15)
100
(3.94)
Mechanical Connection
Collar 22 mm for cutting ring coupling
Parts in contact with fluid
Rod: ceramic, coated
Fluids
1)
Hydraulic oils (mineral-based), synth. oils,
fluids containing water
Temperatur
Measuring range
2)
-25 .. +100 °C
(-13 .. +212°F)
Output variables
Switching outputs
PNP transistor switching outputs
Switching current: max. 250 mA each switching output
(ohmic resistance)
Switching cycles: > 100 million
Analogue output, permitted load
resistance
4 .. 20 mA ohmic resistance max. 500
Ω
0 .. 10 V load resistance min. 1 k
Ω
Accuracy
Level:
≤
± 2 % FS / Temperature: ± 1.5 °C (± 3.0 °F)
Temperature drift (environment)
≤
± 0.015 % FS / °C [0.0085 % FS / °F]
Repeatability
3)
Level:
≤
± 2 % FS / Temperature: ± 1.5 °C (± 3.0 °F)
Reaction time to DIN 60751
(temperature probe)
t
90
̴
180 s
Environmental Conditions
Ambient temperature range
0 .. +60 °C
(32 .. +140°F)
Storage temperature range
-40 .. +80 °C
(-40 .. +176 °F )
Fluid temperature range
0 .. +60 °C
(32 .. +140°F)
Max. tank pressure
0.5 bar (short-term 3 bar, t < 1 min) /
7 psi (short-term 40 psi, t < 1 min)
mark
EN 61000-6-1 / 2 / 3 / 4
mark
4)
Certificate no.: E318391
Vibration resistance acc. to
DIN EN 60068-2-6 (0 .. 500 Hz)
≤
5 g
Shock resistance acc. to
DIN EN 60068-2-27 (1 ms)
≤
25 g
Protection type to DIN EN 60529
5)
IP
67
Other data
Supply voltage
when applied acc. to UL specifications
9 .. 35 V DC, if PIN 2 = SP2
18 .. 35 V DC, if PIN 2 = analogue output
– limited energy – according to 9.3 UL 61010;
Class 2; UL 1310/1585; LPS UL 60950
Current consumption
≤
0.535 A
with active switching outputs
≤
35 mA
with inactive switching outputs
≤
55 mA
with inactive switching output
and
analogue
output
Residual ripple of supply voltage
≤
5 %
Display
4-digit, LED, 7 segment, red, height of digits 7 mm
Weight g
180
220
250
300