6.11 PQA: MEASURING MAINS PARAMETERS IN SINGLE-PHASE SYSTEMS
This function allows measuring
in real time
mains voltage and phase current (with
optional clamp transducer), the corresponding harmonics and evaluating power and power
factor in single-phase systems.
Fig. 38: Connection for measurements in a single-phase system
1. Press the
MENU
key, move the cursor to
PQA
in the
main menu by means of the arrow keys (
,
) and
confirm with
ENTER
.
Subsequently, the instrument
shows a screen similar to the one on the side.
PQA
15/10 – 18:04
V L - N = 0 . 0 V
I =
0 . 0 A
P = 0 k W
Q = 0 k V a r
S = 0 k V A
P f = 1 . 0 0
C o s
= 1 . 0 0
Par 100A
MODE
FS
2. Use the arrow keys
,
to select the parameter to be modified, and keys
,
to
change the parameter's value
:
MODE
This key allows setting the type of display of the parameters measured
by the instrument.
Following options are available:
Par
(mains parameters Voltage,
Current, active, reactive, apparent Power, Power factor, Cos
),
ArmV
(Voltage
harmonics up to the 25th + THDV%),
ArmI
(Current harmonics up to the 25th +
THDI%)
FS
This key allows selecting the full scale
(FS) of the clamp transducers which
can be used with the instrument. Following values are available:
1A, 5A, 10A, 30A,
40A, 100A, 200A, 300A, 400A, 1000A, 2000A, 3000A
3.
Insert the blue and black connectors of the separate cables into the relevant
instrument input terminals B4 and B1. In the free end of the cables, insert the
corresponding alligator clips or leads. Connect the alligator clips, leads to phase P and
to N according to
Fig. 38
to measure voltage in single-phase systems. Connect the
external clamp to the instrument's input
In1
and to the phase conductor for single-
phase systems.
The arrow on the clamp must point in the direction of the current flow,
normally from generator to load, as shown in Fig. 38.