SOLAR300 - SOLAR300N
EN - 138
Energy – Single-phase and Three-phase systems – 50/60Hz
Parameter
[Wh,VArh,VAh]
FS clamp
Range
[Wh,VArh,VAh]
Resolution
[Wh,VArh,VAh]
Accuracy (*)
Active energy
Reactive energy
Apparent energy
FS
1A
0 – 999.9k
0.1 – 0.001k
(0.7%rdg) @
(Pf=1,Vmis>200V,Imis>10%FS)
(0.7%rdg) @
(Pf=1, Vmis > 200V,
Imis>10A if FS=200A)
(0.7%rdg +3dgt) @ (Pf<1 or
Vmis < 200V or Imis < 10%FS
or Imis <10A if FS=200A)
1A< FS
10A
0.000 – 99.99k 0.001k – 0.01k
10A< FS < 200A
0.00 – 999.9k
0.01k – 0.1k
200A< FS
3000A 0.0 – 9.999M
0.1k – 0.01M
(*) For STD clamp ; Vmis = Voltage at which Power is measured
Power Factor (Cos
) – Single phase / Three phase systems
Range
Accuracy (°)
Resolution (°)
0.20
1.00
1.0
0.6
0.01
Voltage / Current harmonics
Range Accuracy(*)
Resolution
DC
25
a
(5.0%rdg+5digits)
0.1V / 0.1A
26
a
33
a
34
a
49
a
(*) Accuracy to add at correspondent TRMS paramters
Frequency
Range Accuracy
Resolution
42.5
69.0Hz
(0.2%rdg+1dgt)
0.1Hz
Flicker – Single phase / Three phase systems
Parameter Range Accuracy
Resolution
Pst1’, Pst
0.0
10.0
compliance to
EN50160
0.1
Plt
Irradiation measurement (by means of remote unit SOLAR-01 and imput PYRA)
Range Accuracy
Resolution Overcharge
protection
0.00
12.00mV
(1.0%rdg+5dgt)
0.01mV
5V
0.00
40.00mV
0.02mV
Temperature measurement (by means of remot unit SOLAR-01 and input TEMP with PT1000 probe)
Range Accuracy
Resolution Overcharge
protection
0
100°C
(1%rdg+2dgt)
1°C 5V