SOLAR300 - SOLAR300N
EN - 120
CAUTION
When SOLAR300N is set in order to use MPP300 as a remote unit, ALL
connections relevant to electrical quantities (voltages and currents) must
be carried out on unit
MPP300
. SOLAR300N must have
no voltage nor
current
connected to its inputs.
The maximum voltage for the inputs of
MPP300
is
1000VDC
between
inputs VDC1, VDC2, VDC3 and
600VAC
between inputs VAC1, VAC2,
VAC3. Do not measure voltages exceeding the limits given in this manual.
Exceeding these limits could result in electrical shocks to the user and
damage to the instrument.
In order to guarantee the operator’s safety, while making the connections,
disable the system being measured by means of the switches/breakers
upstream and downstream of the DC/AC converter (inverter).
1. Check and, if necessary, set the sensitivity of the reference cell on SOLAR-02 consistently
with the type of PV modules which will be measured (please refer to the User Manual of
SOLAR-02).
2. We recommend making a preliminary evaluation of the value of irradiation on the surface of
the PV modules being tested by means of unit SOLAR-02 (operating independently) and the
reference cell. We suggest carrying out the tests only if irradiation is higher than 600W/m
2
3. Switch on SOLAR300N, check and, if necessary, change the settings relevant to the type of
remote unit, to the minimum irradiation threshold, to the full scale of the AC and DC clamps
and to the parameters of the system being measured (see §5.3.3.1, §5.3.3.3, §5.3.1.2,
§5.3.1.3, §5.3.2).
4. In order to guarantee the operator’s safety, disable the system being measured by means of
the switches/breakers upstream and downstream of the DC/AC converter (inverter).
5. Connect SOLAR300N to the MPP300 unit via the USB cable and bring SOLAR-02 and
MPP300 nearer (max distance approx. 1 m).
All instruments must be switched on
(see
the User Manuals of SOLAR-2 and MPP300 for further details). The display of SOLAR300N
must show (for 5 seconds) the message “
MPP300 connected
”.
6. On SOLAR300N, access the
GENERAL MENU
, select the function
View Measures
and
press
ENTER
; wait for the three units to start communicating with each other. This condition
is highlighted by the contemporary presence of the following indicators:
a. Symbol
steady (not flashing) on the display of SOLAR-02
b.
MASTER
and
REMOTE
LEDs flashing green on unit MPP300
7. Fig. 163Fig. 164Connect the
VDC1(+)
and
VDC1(-)
inputs of unit
MPP300
to the output
terminals of the string, respecting the polarities and the colours indicated in Fig. 163 or
Fig. 164.
8. Repeat the operation described in the step above for other possible DC power trackers to be
monitored by using the
VDC2
and
VDC3
inputs according to the number of DC inputs set
(see § 5.3.2.5).
9. Connect the output connector of the DC clamp to the
IDC1
input of unit MPP300