Chapter 4 Command Reference
[:SENSe] Subsystem
4-86
Programming Guide
[:SENSe]:FREQuency:ARM[:STARt]:SOURce IMMediate
|EXTernal
Sets or queries the start arm for Frequency, Frequency Ratio, and Period
measurements.
A sequence of ASCII-encoded bytes: IMM or EXT
*RST: IMMediate
Gate & ExtArm
[:SENSe]:FREQuency:ARM:STOP:DIGits <numeric_value>
Sets or queries the resolution in terms of digits used in arming Frequency, Period,
and Ratio measurements.
3 to 15
Numeric data transferred as ASCII bytes in <NR1> format.
•
*RST: 4
•
Only applies when [:SENS]:FREQ:ARM:STOP:SOUR DIG is selected.
Gate & ExtArm
[:SENSe]:FREQuency:ARM:STOP:SLOPe POSitive | NEGative
Sets or queries the slope of the external stop arm signal used in external arming
Frequency, Frequency Ratio, and Period measurements.
A sequence of ASCII-encoded bytes: POS or NEG
•
*RST: NEGative
•
Only applies when [:SENS]:FREQ:ARM:STOP:SOUR EXT is selected.
Gate & ExtArm
Query Response
Comments
Related
Front-Panel
Keys
<numeric_value>
Range
Query Response
Comments
Related
Front-Panel
Keys
Query Response
Comments
Related
Front-Panel
Keys
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...