Chapter 4 Command Reference
Introduction
Programming Guide
4-81
[:SENSe]:EVENt2:FEED “[:]INPut[1] | [:]INPut2 ”
Sets or queries the common/separate enable.
Feeding the :INPut2 subsystem to the [:SENSe]:EVENt2 subsystem corresponds
to separate; feeding the :INPut1 subsystem to the [:SENSe]:EVENt2 subsystem
corresponds to common.
The string “INP” or “INP2” is returned.
•
*RST: “INPut2”
•
Only the Time Interval function allows
either
the separate or common setting.
All of the other functions have a fixed input route setting.
Trigger/Sensitivity
[:SENSe]:EVENt[1|2]:HYSTeresis:RELative <numeric_value>
[PCT]
The operation of this command is determined, in part, by the firmware revision of
the Counter.
Current Firmware Revision
Sets or queries the size of the hysteresis window as a percentage of the allowable
hysteresis. For example, 0% is the minimum hysteresis setting and 100% is the
maximum hysteresis setting.
Specifying 100% or MAXimum provides the greatest noise immunity (lowest
sensitivity), while specifying 0% or MINimum provides the least noise immunity
(most sensitive).
Prior Firmware Revisions: 3317, 3335, and 3402
Sets or queries the sensitivity of the input channel as a percentage of the allowable
sensitivity. For example, 0% is the minimum sensitivity setting and 100% is the
maximum sensitivity setting.
Specifying 100% or MAXimum provides the most sensitivity
(least noise immunity), while specifying 0% or MINimum provides the lowest
sensitivity (greatest noise immunity).
Query Response
Comments
Related
Front-Panel
Keys
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...