Chapter 4 Command Reference
Introduction
Programming Guide
4-45
•
*RST: OFF
•
When the :INIT:CONT ON command is sent, the Counter:
– invalidates the statistics results,
– clears the statistics current count to 0,
– reports the negative status condition (NOT Computing
Statistics) to bit 8 of Operation Status Register.
•
When :CALC2:LIM:CLE:AUTO is ON, the Counter performs the following
whenever the :INIT:CONT ON command is sent:
– invalidates the limit data,
– clears :CALC2:LIM:FAIL, :CALC2:LIM:FCOunt,
and :CALC2:LIM:PCOunt information,
– turns off the front-panel display’s
Limit
annunciator, and
– sets the Limit-Detect output to the in-limit voltage level.
•
When [:SENS]:EVEN:LEV[:ABS]:AUTO is ON, the Counter performs an auto-
trigger on the measurement channel(s) whenever the :INIT:CONT ON command
is executed, and also at the beginning of each measurement cycle while
:INIT:CONT is ON.
•
The state of :TRIG:COUN:AUTO has no affect on the operation of
:INIT:CONT ON.
•
:INIT:CONT ON operates as if :TRIG:COUN was 1.
•
The commencement of the first measurement due to setting
:INITiate:CONTinuous to ON sets the Pending Operation Flag to true. The
Pending Operation Flag is set false by aborting of a measurement, or by the
completion of the last measurement after :INITiate:CONTinuous is set OFF.
•
With the measurements being made continuously, the :ABORt command shall
abort the current measurement in progress, however, the value of
:INITiate:CONTinuous is unaffected.
If CONTinuous was set to ON prior to receiving :ABORt, it remains ON and
a new measurement begins.
Comments
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...