Chapter 3 Programming Your Universal Counter for Remote Operation
Elements of SCPI Commands
Programming Guide
3-13
Suffix Multipliers
Table 3-2 lists the suffix multipliers that can be used with suffix elements (except
PCT and DEG).
Table 3-2. Suffix Multipliers
DEFINITION
MNEMONIC
NAME
1E18
EX
ETA
1E15
PE
PETA
1E12
T
TERA
1E9
G
GIGA
1E6
MA ( or M for OHM and HZ)*
MEGA
1E3
K
KILO
1E-3
M (except for OHM and HZ)*
MILLI
1E-6
U
MICRO
1E-9
N
NANO
1E-12
P
PICO
1E-15
F
FEMTO
1E-18
A
ATTO
*
The suffix units, MHZ and MOHM, are special cases that should not be confused with
<suffix multiplier>HZ and <suffix multiplier>OHM.
Command Terminator
A command may be terminated with a <new line> (ASCII character decimal 10), an
EOI (End-of-Identify) asserted concurrent with last byte, or an EOI asserted
concurrent with a <new line> as the last byte.
Содержание 53131A
Страница 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Страница 2: ......
Страница 13: ...1 1 Before You Start ...
Страница 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Страница 25: ...2 2 Command Summary A Quick Reference ...
Страница 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Страница 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Страница 167: ...4 4 Command Reference A Dictionary ...
Страница 172: ...Chapter 4 Command Reference CALCulate Subsystems 4 6 Programming Guide Figure 4 1 The CALCulate Subsystems ...
Страница 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Страница 311: ...5 5 Errors ...
Страница 329: ......
Страница 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...